CALIBRATION, RENTAL,TESTING, SALES AND SERVICE




Ultrasonic Flow Meter
Specifications:
Model PORTAFLOW 300:
The Portaflow 300 is a portable flow meter designed for use on liquid flows in full pipes
which utilises “Clamp On” transducers
‘A’ Transducers in mounting rail: 13mm ID to 100mm ID pipes
‘B’ Transducers in mounting rail: 100mm ID to 300mm ID pipes
Guide rail for diagonal operation using ‘B’ or ‘C’ transducers for pipes larger than 300mm
Optional transducers: ‘C’ & ‘D’ for pipes above 300mm upto 6000mm
Transducer operating temperature: ‘A’ ‘B’ ‘C” -25 Deg C to +125 Deg C ‘D’ -25 Deg C to
+70 Deg C. Optional Hi-temp transducers for operation above +125 Deg C to 175 Deg C
Enclosure:
IP66 protection class:
Material: High density P.U. foam
Weight:
Less than 1.5 Kg
Dimensions:
275 x 150 x 55 mm
Display 240 x 64 graphics LCD with backlight
Keypad: IP68 16 key tactile membrane
Connections: IP66 Lemo connectors
Temperature range: 0 Deg C to + 50 Deg C (-10 to 50 Deg C storage)
Supply voltage:
Power supply / charger
Input: 100-260 VAC +/- 10% @ 50/60 Hz Max. 9 Watts
Output: 9VDC unregulated
Battery pack:
Internal batteries
5 x 4/3 AA nickel metal hydride: 24-30 hrs continuous operating on fully charged battery
Recharge time: 10-16 hours
Outputs: Display
Volumetric flow units: m3, litres, gallons (imperial/ US)
Velocity units: metres/sec, feet/sec
Flow velocity range: 0.2…12 m/sec to 4 significant figures
Total volume: 12 digits – forward & reverse
Analogue: 4-20mA into 750 ohm user definable scaling
Resolution 0.1% of fullscale
Pulse: 5 volts Max. 1 pulse per second
Data Logger: Memory capacity: 100K (50000 readings)
Output: Via RS232 or displayed graphically
Logs: Block data storage with test & graphic display, transferred to Microsoft windows
Transducer sets:
‘A’ (standard) 13mm to 115mm pipe 0.2m/sec to 8 m/sec
‘B’ (standard) 50mm to 1000mm pipe 0.2m/sec to 12 m/sec
‘C’ (optional) 300mm to 2000mm pipe 0.2m/sec to 7 m/sec
‘D’ (optional) 1000mm to 5000mm pipe 0.2m/sec to 7 m/sec
Temperature range ‘A’ ‘B’ ‘C’ -20 to +200 Deg C standard
Frequency: 1MHz, 2 MHz, 0.5 MHz
Pipe Material:
Any sonic conducting medium such as crbon steel, stainless steel, copper, UPVC, PVDF,
Concrete, Galvanised steel, Mild steel, Glass, Brass, including lined pipes-Epoxy, Rubber,
Steel, Plastic
Repeatability:
+/- 0.5% with unchqnged transducer position
Accuracy:
2% or +/- 0.02 m/sec which ever is greater. Accuracy achieved under ideal calibration conditions on a 4” plastic pipe. Specifications assumes turbulent flow profile with Reynolds
numbers above 4000








Click on image to enlarge
-